Description
The new NeoFox Phase Measurement system is a fluorescence-based optical sensor system that helps you reduce costs, improve stability and make calibration easier for a wide variety of oxygen sensing setups. NeoFox is the benchtop device for measurement of fluorescence lifetime, phase and intensity and is particularly suited for applications where sensitivity to drift and system stability are critical.
Plus, NeoFox is about half the cost of comparable phase measurement systems.
|
FOXY Formulation
(general-purpose coating) |
FOSPOR Formulation
(high-sensitivity for low levels of O2) |
HIOXY Formulation
(for hydrocarbon environments) |
O2% range: |
0-100% |
0-20.9% |
0-100% at 1 ATM |
DO range (ppm): |
0-40 ppm |
0-8 ppm |
0-40 ppm |
Temperature range: |
-50-+80 ºC for probes |
-50-+80 ºC for probes |
-50-+60 ºC |
0-60 ºC for patches |
0-60 ºC for patches |
NA |
O2% resolution: |
0.01-0.05% |
.001% |
0.01-0.05% |
DO resolution: |
4-20 ppb at room temp. |
0.4-2 ppb at room temp. |
4-20 ppb at room temp. |
O2% accuracy: |
5% of reading |
5% of reading |
5% of reading |
DO accuracy: |
5% of reading |
5% of reading |
5% of reading |
Min. detectable level: |
100-500 ppm |
10-100 ppm |
100-500 ppm |
Response time: |
< 1 second in gas |
< 1 second in gas |
< 1 second in gas |
~30 seconds with overcoating in gas |
~30 seconds with overcoating in gas |
~45 seconds in pure water |
|
~45 seconds in pure water |
~45 seconds in pure water |
|
Sterilization Guide:
Coating |
Form factor |
Autoclav-
able |
Calibration |
Comment |
HIOXY |
Probes |
Yes |
Single point reset required
after sterilization |
Can be sterilized multiple times |
|
Custom RedEye patches |
Yes |
New calibration required
after sterilisation |
Can be sterilized a maximum of 2 times |
FOXY |
Probes/Patches |
No |
|
|
FOSPOR |
Probes/Patches |
No |
|